Anritsu ML2430A Manuel d'utilisateur Page 119

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 271
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 118
Remarks:
Sets the mode of the LCD backlight when under Battery power.
ON = back light is ON all the time
OFF = back light is OFF all the time
TIMED = back light is on for a limited time period set by the DBLTIM com
-
mand.
NOTE
Although GPIB is not available under battery operation, the
state of this battery-specific parameter can be changed
through this GPIB command.
Related
Commands:
DBLTIM
Query:
DBLGHT?
Returned
String:
DBLGHT <mode>
DBLTIM Auto Backlight OFF timer setting
SYSTEM
Syntax:
DBLTIM <val>
val:
1.0 to 100.0 minutes
Remarks:
Sets the time limit when the backlight will turn off if the DBLGHT setting is
set to TIMED.
NOTE
Although GPIB is not available under battery operation, the
state of this battery-specific parameter can be changed
through this GPIB command.
Query:
DBLTIM?
Returned
String:
DBLTIM <val>
DCONT Set Display Contrast
DISPLAY
Syntax:
DCONT <val>
val:
1 to 10
Remarks:
One is the lightest setting, ten the darkest. The default is five.
Query:
DCONT?
6-32 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION
Vue de la page 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 270 271

Commentaires sur ces manuels

Pas de commentaire