SECTION 6 GPIB
6-41
6.7.5 PM subsystem
:PM:STATe <Boolean>
Function Turns ON or OFF the pulse modulation.
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:PM:STATe?
Function Reads out the state of pulse modulation.
Response Pulse modulation = OFF
ON
Restriction None
:PM:IMPedance <Character>
Function Selects the impedance of pulse modulation input.
Parameter <Character> = HIGH (600 )
LOW (50 )
Unit <Non term>
Restriction None
:PM:IMPedance?
Function Reads out the state of impedance of pulse modulation input.
Response Impedance of pulse modulation input = HIGH
LOW
Restriction None