To set the detection method, press the DETECTION soft key and select either POSITIVE
PEAK, AVERAGE,
or NEGATIVE PEAK detection.
Sweep Average
In order to reduce the effects of noise, it may be desirable to average the results of several
sweeps and display that average rather than the results of individual sweeps.
To set the number of sweeps to average, press the
AVERAGE (1-25) soft key and use the
up/down arrow key or directly enter the desired number of sweeps from the keypad. Press
ENTER to set the sweep average.
NOTE: MAX HOLD and AVERAGE are mutually exclusive.
Adjusting Attenuator Settings
The spectrum analyzer attenuation can be automatically coupled, manually coupled, or dy
-
namically adjusted.
Step 1.
Press the AMPLITUDE key.
Step 2. Press the
ATTEN soft key.
Step 3. Select the soft key corresponding to the desired coupling mode, as described be-
low.
Auto Coupling
Auto coupling of the attenuation links the attenuation to the reference level. That is, the
higher the reference level, the higher the attenuation. Auto coupling is indicated on the dis-
play as
ATTEN XX dB.
Manual Coupling
When manually coupled, the attenuation can be adjusted independently of the reference
level. Manual attenuation coupling is indicated on the display as
ATTEN* XX dB.
IMPORTANT!
The attenuation should be adjusted such that the maximum signal amplitude at
the input mixer is –30 dBm or less. For example, if the reference level is +20
dBm, the attenuation should be 50 dB such that the input signal at the mixer is
–30 dBm (+20 – 50 = –30). This prevents signal compression.
Dynamic Attenuation
Dynamic attenuation tracks the input signal level, automatically adjusting the reference
level to the peak input signal. When dynamic attenuation is enabled, the attenuation is auto
-
matically coupled to the reference level.
Dynamic attenuation is indicated on the display as
ATTEN# XX dB.
3-11
Chapter 3 Getting Started
Vue de la page 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 119 120

Commentaires sur ces manuels

Pas de commentaire

Behringer V-AMP2 manuali

Manuali dei proprietari e guide per l'utente per Lettori DVD Behringer V-AMP2.
We fornisce manuali pdf Behringer V-AMP2 per scaricare gratuitamente in base al tipo di documento:






Più prodotti e manuali per Lettori DVD Behringer

Modelli Tipo di Documento
Nessuno