4-8 S-PARAMETERS
CAL
The “C_S-Parameters Cal.txt” file to provides prompts for setting up
S-parameters specification file. The prompts cause the Scorpion to be
placed into a state where it is ready to make calibrated measurements on
the users AUT. At the completion of the program, the state information is
saved to a *.cal file. To run this program, proceed as follows.
Step 1. Select “Tools” then “Run Calibration Procedure.”
Step 2. Double-click on “D_S Parameters.txt” to begin the file
setup.Follow the prompt and connect open, short, and load to
Ports 1 and 2 as directed.
OPERATIONS, CALIBRATION S-PARAMETERS CAL
ME7840A OMM 4-11