Step 3. Press “Next” for the next step (below). The 3653 or 3753LF
Calibration Kits contain a disk with calibration data. The
data has to be installed from the MS462C front panel. To do
so, place the floppy disk from the kit into the MS462XC disk
drive. Press the Cal key then COMPONENTS UTILITIES/IN
-
STALL KIT INFO FROM FLOPPY DISK softkeys.
Step 4. Press “OK” for the next step (below). Connect the Broadband
Load on Test Ports 1 and 2 on the ME7840A Test Set.
Step 5. Press “OK” for the next step (below). Connect the Open to
Test Port1 and the Short to Test Port 2.
S-PARAMETERS CAL OPERATIONS, CALIBRATION
4-12 ME 7840A OMM