4-9 CAL FILE SETUP The “D_Cal File Setup.txt” file provides prompts for setting up One-tone
power sweep, IMD, Two-Tone Power Sweep, and Harmonic specification
file. The prompts cause the Scorpion to be placed into a state where it is
ready to make calibrated measurements on the users AUT. At the com
-
pletion of the program, the state information is saved to a *.cal file. To
run this program, proceed as follows.
Step 1. Select “Tools” then “Run Calibration Procedure.”
Step 2. Double-click on “B_Cal Files Setup.txt.” to begin with the file
setup for a Transmission Frequency Response (S
21
cal) for
One-Tone Power Sweep. Connect a throughline when di
-
rected in a following step.
OPERATIONS, CALIBRATION CAL FILE SETUP
ME7840A OMM 4-15